Date of Issue: 12 September 2024
IFU Ref: REC-QMS-053
Revision:  6
If revised date of revision: 12 September 2024
Change Control Number: CC-2024-001
Name of device: Lifelight® EA
Trademark: Lifelight®

factory Manufacturer

XIM Ltd.
University of Southampton Science Park
2 Venture Road
Southampton
Hampshire
United Kingdom
SO16 7NP

ll-qr

(01)05060776720026
(11)240912
(21)LLEA-001
(8012)6.0.0-1

Symbols on the device and documentation

The symbols and labels that appear in this manual and within the application have the following meanings:

Symbol Description
factory

Serial Number.
The serial number for this device is: LLEA-001.

factory

Catalogue Number.
The Catalogue number for this device is: LLEA-001.

factory

Consult Instructions For Use.

factory

Indicates the item is a medical device.

factory

Date of manufacture.

factory

Name and address of the manufacturer.

caution

Caution.

ce

Complies with the European Medical Device Directive (93/42/EEC) and amendments by directive 2007/47/EC.

ce

Complies with the UK Medical Device Regulations 2002 (UK MDR).

ce

Name and address of the Authorised Representative in the European Union (EU)

ce

Name and address of the Authorised Representative in Switzerland

ce

Name and address of the Importer for the specified country/region/jurisdiction

Lifelight® EA  Intended Use / Indications For Use

Lifelight® EA non-contact medical device software is intended for indicative estimates for:

  • non-invasive blood pressure (mmhg),
  • pulse rate (bpm),
  • and respiration rate (rpm)

Intended patient populations include: adult patients. The medical device software is not intended for paediatric use.

The most likely locations for use are hospitals, clinics, residential care, other healthcare environments and the patient’s home.

The Lifelight® EA medical device software is to be used either by care providers, or by a patient who has been advised in its appropriate use by a care provider.

Lifelight® EA is indicated for use with patients at rest, during no-motion, in a well-lit environment.

NOTE: Lifelight® EA is intended for informational use only. It should not be used for direct diagnosis or monitoring of vital physiological processes and is not intended to replace traditional methods of diagnosis or treatment.

Exceptions / Warning:

For the attention of healthcare practitioners:
Some drugs, therapies, health and skin conditions (including cosmetics, facial tattoos and excessive sweat) may interfere with readings – therefore clinicians must verify all vital signs information before using results from Lifelight® EA for informational purposes.

If there is any reason to question the accuracy of the results, use other clinically valid methods for verification.

For the attention of patients and/or their carers and care providers:
Please be aware that vital signs can fluctuate, since they respond to small changes in the body, and with the time of day. It is unusual for them to remain at exactly the same value all the time. If you are checking your vital signs regularly, e.g. blood pressure on different days, you should ideally take readings at the same time and under the same conditions.

If a reading returns results that seems to be abnormal, but you do not feel unwell, do not panic, as this can happen. Relax and take a break for 5 minutes before checking vital signs again. Low blood pressure (hypotension), high blood pressure (hypertension), the reasons for consistently fast (tachycardia) or consistently slow (bradycardia) heart rate; or for unusual respiratory rates, can only be diagnosed by a qualified healthcare professional. If you have any concerns about your results, please contact your GP surgery or pharmacist.

Ideally, to get the most accurate values, you should wait at least an hour after eating, smoking, caffeine, alcohol, or exercise, as these may affect the results. You should be relaxed and keep still during the reading. Do not use this medical device software if you suffer from atrial or ventricular premature beats or atrial fibrillation.

Terminology

Warnings and Statements that can appear in the app

Warning and statements can appear within the app or in this document. When a warning is issued the warning triangle below will be displayed.

caution

The app is safe for patients and clinicians when used in accordance with the instructions and the warning statements presented in these Instructions For Use.

Before using the app, you must familiarize yourself with all warnings to ensure good reading quality and usable readings.

In addition to reviewing the general warnings presented in the next section, you must also review the more specific instructions found in the Preparation and Set-up, Taking a Reading and Maintenance sections.

  • Failure to understand and observe any warning statement in this manual could lead to poor reading quality and no readings returned.

caution Low Reading Quality

Low reading quality, leading to no results returned at the end of the reading, can occur with one or more of the following causes:

Lighting

  1. The subject (the person being checked) is poorly lit.
  2. The subject is strongly lit from behind or above and not adequately from in front.
  3. The background to the subject is very brightly lit.
  4. A mirror or reflective surface is in the background of the subject.
  5. There are strong shadows on the subject’s forehead and cheeks.
  6. The light changes during the reading. Other than flickering lights, this could be caused by people or objects moving in the background.

Movement

  1. The device is moving too much during the reading. If you can't hold the device still, avoid hand-holding the device and instead use a stand or steady the device against an object.
  2. The subject may be moving too much such as talking or looking around during the reading. (Blinking and slight movements caused by breathing should not affect reading quality.)

Other

  1. There is not an adequate wifi or data connection to upload the data.
  2. The subject does not have their forehead and cheeks visible.
  3. The device lens is not clean.

What Lifelight® EA does and how it works

Lifelight® EA is a software application that is a medical device. It is intended to be a simple way to estimate pulse, respiration and blood pressure in 40 seconds or less. Through the device camera, the software measures changes in facial skin colour each time the heart beats. When the colour information is processed, the software provides indicative estimates for pulse, respiration and blood pressure.

Preparation and set up of Lifelight® EA before taking a reading.

Device preparation and minimum requirements.

IT networks characteristics and IT security measures.

Keeping the device still.

Environment, lighting and special facilities.

Subject – the person being checked.

Taking a Reading

Ensure the preparation and set up steps described previously have been followed.

app icon Lifelight® EA app icon. Start the application by clicking this icon.

Enter your biometric information - sex, age and height

Taking a Reading

Reading Quality Indicators

When to retake a reading

You can retake a reading if there is any reason to assume a poor reading has occurred.

It is possible to cancel the reading before the end and restart.

Other

Maintenance / Frequency

Update software and device operating system when updates become available.

Clean device when necessary.

Serious Incident

Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or subject is established.

Troubleshooting / Precautions if malfunction

Situation Mitigation

The application freezes.

Results will not be returned at the end of a reading. Quit and restart the application or host device.

Unable to use because no data connection or wi-fi signal.

Find or wait for signal and restart. Or use alternative conventional methods.

Camera not giving clear image due to dirt or condensation.

Clean the lens.

Mains recharging not available.

Have backup power pack available to recharge the device.

Camera not giving clear image due to dirt or condensation.

Use alternative conventional methods.

Before using this product, the healthcare provider should study carefully the indications, intended use, recommendations, warnings and instructions, as well as the available product-specific information. Xim Ltd is not liable for problems arising from the use of the device outside of its indicated uses and similar matters outside the control of xim Ltd.

Information provided in this Instructions for Use is intended for care providers, or by a patient who has been advised in its appropriate use by a care provider..

For any question or further information on this instructions for use, please make contact via www.lifelight.ai

factory

ARAZY GROUP [IRELAND] LIMITED
19 Baggot Street Lower,
Dublin 2
D02 X658
Ireland
ireland@arazygroup.com

factory

MedEnvoy Switzerland
Gotthardstrasse 28
6302 Zug
Switzerland

factory

CH Importer
Sanitas
Sanitas Management AG,
Jägergasse 3,
8004
Zürich
Switzerland